Saturday, May 24, 2008

New publication: Full-range, high-speed 1um SD-OCT

Full-range, high-speed, high-resolution 1μm spectral-domain optical coherence tomography using BM-scan for volumetric imaging of the human posterior eye

This our paper is now available on Optics Express.

>>[Abstract and PDF]

Citation
S. Makita, T. Fabritius, and Y. Yasuno, "Full-range, high-speed, high-resolution 1μm spectral-domain optical coherence tomography using BM-scan for volumetric imaging of the human posterior eye," Opt. Express 16, 8406-8420 (2008).


Abstract
An alternative optical coherence tomography (OCT) to clinical ophthalmic 830 nm spectral-domain OCTs (SD-OCT) is demonstrated. An axial resolution of 7.4 μm, ranging depth of 4.2 mm in tissue, sensitivity of 98.5 dB, and detection speed of 38,300 axial scans/s have been achieved. These are comparable or superior to those of recently commercially available ophthalmic 830 nm SD-OCTs in clinics. In addition, fast volumetric imaging for the in vivo human posterior eye with high-contrast of the choroid is achieved. A broadband 1.04 μm light source enables the high-contrast and high resolution imaging of the retina and choroid. The ranging depth is extended by applying a full-range imaging method with an electro-optic modulator (BM-scan method). A prototype high-speed InGaAs line scan camera with 1024 pixels is used. A newly reported sensitivity improvement property of the BM-scan method demonstrates a sensitivity enhancement of 5.1 dB. We also introduce a newly developed resampling calibration method of spectrum that is independent of the intrinsic dispersion mismatch of the interferometer. The three-dimensional structure of the in vivo human optic nerve head with a very deep cupping is successfully visualized.